ABSTRACT
BACKGROUND/AIMS: A barbed puncture needle with a side hole was recently developed to improve sample quality and quantity in endoscopic ultrasound-guided fine-needle aspiration (EUS-FNA). In this study, we retrospectively assessed the usefulness of this puncture needle. METHODS: Factors affecting diagnostic yield, safety, and diagnostic accuracy were investigated in 76 patients who consecutively underwent EUS-FNA for neoplastic lesions at our hospital between January and December 2013. RESULTS: The procedure was successful in all cases; the rates of sample collection and determination of the correct diagnosis were 92.1% and 89.5%, respectively. The mean number of needle passes required for diagnosis was 1.1. Complications included mild intraluminal bleeding in two patients (2.6%). Multivariate analysis revealed that lesion size (< or =20 mm) was significantly associated with a decreased chance of determining the correct diagnosis. CONCLUSIONS: Core biopsy needles with a side hole are safe and provide a satisfactory diagnostic yield. However, the side hole may potentially reduce the rate of making the correct diagnosis in small lesions.